Technical characteristics:
- Temperature and Humidity control.
- Temperature range: +5°C/+70°C (accuracy ±1°C all over inside the chamber) (read on display ± 0,1).
- Humidity range: 30% to 90% (accuracy ±2 all over inside the chamber) (read on display ±0,1).
- Construction in stainless steel 18/8 as per AISI 304 both inner and outer case.
- High density insulation through foamed-in-place polyurethane.
- Stainless steel frontal door with lock and key.
- Transparent inner glass door.
- Refrigerating Device located in the upper part of the chamber with C.F.C. free air cooled device, inner ventilation to uniform temperature.
- Heating through Resistors easy to be replaced.
- ∅ 35 Hole/Port.
- Humidity Device through ‘Carel’ type UR hot steam Generator to be filled with pressured treated water (in case laboratory is not provided with pressured treated water, a tank with a pump will be provided).
- Humidifier positioned and fixed outside, behind the chamber.
- De-Humidification Device.
- Defrosting System.
Control Panel Board completed with:
- Temperature Digital Microprocessor Thermo-Regulator Ascon M33107, auto-tuning P.I.D. control 4-20 mA signal re-transmission for Recorder connection.
- Humidity Digital Microprocessor Thermo-Regulator Ascon M33107, auto-tuning P.I.D. control 4-20 mA signal re-transmission for Recorder connection.
- Min/max safety thermostats with visible alarms.
- Min/max Temperature Alarm.
- Min/max Humidity Alarm.
- Safety Lamps.
- Humidity Device Switch.
- Mains Switch.
- Fuses.
V. 230/1 Hz. 50
CEI 62-25 Electric device – CE Marking
Mod. 900CFUD
Capacity lt.720
- Int. Dim.: mm.640 x 750 x 1500H.
- Ext.Dim.: mm.790 x 1300 x 2070H.
- N° 4 Shelves (grid easy to move).
- N° 4 Wheels/Castors Kit (n.2 with stop block).
Options:
- Temperature/Humidity Protocols.
- IQ/OQ Protocols and Protocol Tests.
- Video Recorder with colour Display ‘Memograph.
- M RSG40-B111B1C1A1A1 with n.4 channels.
- FDA 21CFR Part 11 User administration.